Product Introduction
	            GJX-3 Vibration Calibrator(Metal Frame Case)
 
This Vibration Calibrator can be used to calibrate vibration accelerometer, velocity and proximity transducer and also a vibration measuring instruments. Standard sine signals at frequency 10, 20, 40, 80, 160, 320, 640 and 1280 Hz can be generated from this Vibration Calibrator. The amplitude of the, acceleration, velocity and displacement, can be defined through potentiometers and shown on digital displays. Either horizontal or vertical vibration can be generated by this Vibration Calibrator.
GJX-3 Vibration Calibrator combines sine signal generator, power amplifier, standard transducer and a shaker in one and features a small volume, high accuracy and easy operation. It can be used either in laboratory or on site.Combined with a portable power bank, it can work outdoors for more than 8 hours.
Vibration Calibrator Specifications
Frequency: 10, 20, 40, 80, 160, 320, 640, 1280Hz +0.01%
Amplitude accuracy:
Acceleration(@30m/s2 pk) 40Hz to 320Hz + 0.3dB + 1digit
20Hz to 640Hz + 0.5dB +1digit
Velocity (@25mm/sec pk) 40Hz to 320Hz +0.5dB +1digit
Displacement(@50μm pk-pk) 40Hz to 320Hz + 0.5dB + 1digit
	
Vibration Calibrator Proximity probe linearity 
Prode: 5mm and 8mm probes Range: 0~4.0mm
	
Vibration Calibrator Display:  3 1/2 digit display for acceleration, velocity or displacement 
	
 
Maximum vibration amplitude and maximum load
| 
				 Wt 
 Freq  | 
			
				 <100g  | 
			
				 <250g  | 
			
				 <650g  | 
		||||||
| 
				 a(m/S2)  | 
			
				 v(mm/S)  | 
			
				 d(μm)  | 
			
				 a(m/S2)  | 
			
				 v(mm/S)  | 
			
				 d(μm)  | 
			
				 a(m/S2)  | 
			
				 v(mm/S)  | 
			
				 d(μm)  | 
		|
| 
				 10Hz  | 
			
				 2.5  | 
			
				 28  | 
			
				 1300  | 
			
				 3.5  | 
			
				 40  | 
			
				 1800  | 
			
				 4  | 
			
				 45  | 
			
				 2000  | 
		
| 
				 20Hz  | 
			
				 15  | 
			
				 85  | 
			
				 1900  | 
			
				 10  | 
			
				 60  | 
			
				 1300  | 
			
				 5  | 
			
				 28  | 
			
				 640  | 
		
| 
				 40Hz  | 
			
				 60  | 
			
				 170  | 
			
				 2000  | 
			
				 35  | 
			
				 100  | 
			
				 1100  | 
			
				 12  | 
			
				 35  | 
			
				 380  | 
		
| 
				 80Hz  | 
			
				 100  | 
			
				 141  | 
			
				 800  | 
			
				 40  | 
			
				 60  | 
			
				 320  | 
			
				 14  | 
			
				 20  | 
			
				 110  | 
		
| 
				 160Hz  | 
			
				 75  | 
			
				 53  | 
			
				 150  | 
			
				 35  | 
			
				 25  | 
			
				 70  | 
			
				 12  | 
			
				 8.5  | 
			
				 24  | 
		
| 
				 320Hz  | 
			
				 50  | 
			
				 18  | 
			
				 25  | 
			
				 30  | 
			
				 10  | 
			
				 15  | 
			
				 10  | 
			
				 3.5  | 
			
				 5  | 
		
| 
				 640Hz  | 
			
				 30  | 
			
				 5  | 
			
				 3  | 
			
				 20  | 
			
				 3.5  | 
			
				 2  | 
			
				 6  | 
			
				 1  | 
			
				 *  | 
		
| 
				 1280Hz  | 
			
				 23  | 
			
				 2  | 
			
				 *  | 
			
				 10  | 
			
				 0.9  | 
			
				 *  | 
			
				 5  | 
			
				 0.4  | 
			
				 *  | 
		
	
Size:  280 x 180 x 250(mm)                       
Weight:  10kg 
	Temperature: Operation:  0oC to 50 oC   
Storage:  -20 oC to 70 oC 
Humidity: 90% non -condensing
The control panel of GJX-3 Vibration Calibrator is shown as following:
	
 
1 Vibration amplitude output display
Showing the amplitude of the different vibration signal output of the calibrator
2 Calibration potentiometer
3 Function selecting switch Shifting the output of the calibrator to acceleration, velocity or displacement
4 Potentiometer for adjusting the output amplitude
Adjusting the potentiometer to change the output vibration amplitude
5 Adjusting the the output vibration amplitude slowly
6 The table for fixing the transducer to be calibrated
7 The thread hole for fixing the stand of proximity probe
During the calibration of proximity probe, the stand of the transducer is fixed on
the control panel through this thread hole.
8 Power switch 220 VAC
9 Socket for 220 V AC power input
Used for connecting 220 V AC power.
10. Socket for power output
For monitoring the wave shape of the output signal of the power amplifier of the
calibrator.
11 Output socket of – 24 V voltage
Providing power at – 24 V for proximitor during the calibration of proximity probe.
12 Frequency selecting switch
For selecting the frequency of the output signal
	
	
 
	
 
	
	 
						 
						 
						 
						 
						
		
			
 
				 
			
					 
			
					
				 
		
	
					 
				
					
					
					
					
					 
			
				
	
 
	


                        







                              
                              
                              
                              